MAG*I*CAL®
Calibration Reference Standard for TEM
The MAG*I*CAL®
was developed in a world-class Canadian Solid
State Physics laboratory to perform the three major
calibrations for transmission electron microscopy.
-
The
image magnification calibration – All TEM
magnification ranges
-
The
camera constant calibrations (indexing diffraction
patterns)
-
The
image/diffraction pattern rotation calibration
-
Directly
traceable to a natural constant
MAG*I*CAL® is
the world’s smallest ruler and has been inducted into
The Guinness Book of World Records.
The MAG*I*CAL® offers real
advantages over other calibration samples:
-
It is directly
traceable to the crystal lattice constant of silicon.
This constant can be measured directly on the MAG*I*CAL®
sample, providing unbroken traceability to a fundamental
constant of nature.
-
Instead of 4-6
calibration samples to cover all the magnification
ranges on a TEM, the MAG*I*CAL® calibration standard
covers all ranges with one sample, offering unparalleled
accuracy and consistency.
-
The MAG*I*CAL®
can also be used for the electron diffraction (camera
constant) calibration, as well as the calibration of the
rotation between the image and diffraction pattern.
-
No other
calibration sample can rival the accuracy, reliability
and traceability of the MAG*I*CAL®.
Magnification Calibration
Standard for TEM

MAG*I*CAL® calibration
reference standard for TEM consists of an ion-milled
cross-section of a silicon wafer on which a series of
calibration marks were deposited. The spacing between these
calibration marks are very accurately known. When the
calibration structure is viewed in a TEM, it appears as a
series of light (silicon) and dark layers (SiGe).The
calibration thickness measurements of the light and dark
layers are based on careful TEM measurements of the <111>
lattice spacing of silicone, which is visible on the
calibration sample itself and are supported by X-ray
diffraction measurements. The layer spacing are designed so
the sample can be used to calibrate the entire magnification
range in TEM from 1,000X to 1,000,000X.
As the sample is also a single crystal of
silicone, the calibrations requiring electron diffraction
information, such as the camera constant and
image/diffraction pattern rotation can also be performed
easily.
One single calibration sample can
therefore be used to provide all three of the major TEM
instrument calibrations at all magnifications and all camera
lengths
Diagram of the MAG*I*CAL®
Grid.


The arrows on the drawing
and photograph indicate the four regions on the standard
where the calibration marks may be found.
canliv@canemco.com 1-877-633-0550